The Power Puck II UV Radiometer is a four channel radiometer that measures peak irradiance (W/cm2) and energy (J/cm2) on a production line, and has the ability to measure and display the irradiance profile information directly on the radiometer making it easier to detect poorly performing UV lamps in a multi-lamp system. A 10 watt (high) system (050–60530), the Power Puck II also comes in a 1 watt (medium) system (050-60530-M) and a 100 mW (low) system (050-60530-L).
With the added ability to download data for viewing, comparison, analysis and archiving using the UV PowerView Software III, the UV Power Puck II Profiler is available in a 1 watt (050-60537-M) Profiler, 10 watt (050-60537) Profiler, and 100 mW (050-60537-L) Profiler.
Display: Easy to Read, Yellow Text on Black Background
User Selectable Dynamic Operating Ranges are: Standard High Range is UVA, UVB, UVV at 100mW/cm² – 10W/cm² and UVC at 10mW/cm² – 1W/cm² Mid-Range is UVA, UVB, UVV at 10mW/cm2 – 1W/cm2 and UVC at 1mW/cm2 -100mW /cm2 Low Power is UVA, UVB, UVV at 1mW/cm² – 100mW/cm² and UVC at 1mw/cm2 to 100mW/cm2 Units will “turn on” and display data at irradiance values much lower than the suggested operating ranges. The suggested operating ranges are where the instrument performs best.
Accuracy: +/- 10%; +/- 5% typical
Spectral Ranges and channels for the UV Power Puck II are as follows: The UV Power Puck® II is a four channel continuous monitoring radiometer. The spectral ranges for the Standard Version are 320-390nm UVA, 280-320nm UVB, 250-260nm UVC, and 395-445nm UVV. The UVA2 Version replaces UVC with UVA 2 and the spectral range is 380-410nm. UVA, UVB, and UVV ranges remain the same.
Spatial Response: Approximately cosine
Operating Temperature: 0-75°C internal temperature and tolerates high external temperatures for short periods. An audible alarm indicates when temperature has exceeded tolerance.
Time-Out Period: 2 minutes DISPLAY mode with no key activity. A no time-out mode can be ordered.
Batteries: Two user-replaceable AAA Alkaline Cells
Battery Life: Approximately 20 hours with display on
Dimensions: 4.60 x 0.50 inches; 117 mm x 12.7 mm (D x H) Weight: 10.1 ounces (289 grams)
Instrument Materials: Aluminum and stainless steel.
Carrying Case Material: Cut polyurethane interior with a scuff resistant nylon exterior cover.
Carrying Case Weight: 9 ounces (260 grams)
Carrying Case Dimensions: 10.75 x 3.5 x 7.75 inches; 274 x 89 x 197 mm (W x H x D)
146 Log Canoe Circle
Stevensville, Maryland 21666 USA
Phone: +1 410-604-2900
Email: service@miltec.com
+1 443-591-2326
Phone: +1 410-604-2900
Fax: (+1) 410-604-2906
Email: sales@miltec.com
Photography by David Bohrer, National Assoc. of Manufacturers and Gary Landsman, Gary Landsman Photography
©2024 Miltec UV. All Rights Reserved. Privacy Policy / Terms and Conditions